Back to
Instrumentation List
|
Level Control Setup
- BSPIL-IST-05025
Level Control Setup comprises of Measuring tank fitted with Level transmitter to
measure the Water Level. The Flow of water is controlled by a control valve
actuated by a I to P converter. Rotameter is provided to measure the flow rate.
The system is mounted on a table top model with all necessary pipings. The water
inlet and outlet is provided through a manual control valve. Water at approx.
100 LPH has to be supplied to the system externally. The electronic panel
comprises of ADC/DAC card with RS-232 interface. Displays indicate the process
value and set value. Software is provided for experimentation, PID Control
settings (P, PI, PD and PID mode) Data storing, Online data acquisition, Off
line analysis, Printing of data's can be done easily.
Specification :
Level Sensor : Capacitance Type Level Sensor.
Output : 4 - 20m A
Measuring Tank: Capacity 5 liters, with graduation.
Rotameter : 10-100 LPH
Control Value :Globe 2 way ½" with Diaphragm type actuator, Total stem travel
18mm, Air operated, 3 to 15 PSI pressure.
Air Regulator : ¼" BSP, for 1 to 15 PSI pressure
I/P Converter :4 to 20mA input output 3 to 15 PSI
ADC/DAC Card : 12 bit resolution, 8 input 1 output with RS 232 Serial Interface
to PC.
PID Controller :Software to set P, PI, PID, PD Mode with Controller and Data
Acquisition, Trend plot, Offline, Analysis and Printing.
|